Quick Guide to using the sample bag

sleep•safe harmony Quick Guide to using the sample bag

 


Follow the instructions

1 Prepare sleep•safe harmony as usual up to the connect drain line screen:

2 Add the sample bag to the fluid circuit by connecting the yellow connectors; 2 & 4 and 1 & 3.

3 Close Clamps 5, 6 & 7, leaving clamp 9 open for prime as usual.

4 When you are ready to collect sample during a drain or outflow cycle, open the small clamp on sample bag (clamp 5) and close clamp on large bag (clamp 9).

5 When the sample bag is full, close clamp on sample bag (clamp 5) and re-open clamp 9 on the large drainage bag to complete the drain outflow cycle.

6 At the end of treatment, disconnect sample bag from circuit and retain as unit instructions.

 